MORECHEMISTRY.COM   MORE CHEMISTRY ON THE WEB

XPS Info - Issues

This page describes some of the technical issues we've run into and how they were solved:

There are at least two possible reasons why the autoheight function might fail.

One reason is a bit trivial. In that case the stage runs into one of its limits. Setting the right starting point and range of operation will avoid this situation.

The other reason is more serious. In that case the stage moves up and down exactly as expected, but after the final step the program stops (with an error). The most probable cause is that the communication with the detector fails, in which case the assistance of the technician (to reset the XPS) is required.

If one suspects that the top camera in the analysis chamber is misaligned, e.g. that the point of measurement doesn't coincide with the location of the X-ray spot, then there is a simple trick to establish this.

On the top-left of the 'Optical View' page select the pre defined position 'Phosphor'. Press the arrow to move there. Turn the internal lights off. Open the X-ray control dialog from the diagram on the 'Sample' page. Turn on the X-ray gun. Allmost immediately the X-ray spot will light up.

If the X-ray spot isn't at the center of the image then the camera is misaligned. Until this has been solved by the service engineer, the XPS can still be used but one should take into account that there is an offset (either horizontally or vertically or in both directions).

If one of the cameras is off, or rather not working, then, in most of the cases, the quickest way to turn it back on is applying the following procedure:

  • close the Avantage software
  • run the server stopper program
  • run the search device program and perform a search for devices
  • start the Avantage program
  • check the result

If Avantage reports that the sampleblock isn't on the stage, then there may be a serious problem. If this happens for the first time, then it can be tried to place the sampleblock back into the park position after which the sampleblock can be moved onto the stage again. If the error persists the help of the service engineer may be required. In any case report the error to the technician.

The vacuum in the loadlock should be below 2.10-7 mbar before a sample can be moved into the analysis chamber. There can be several reasons why this level can't be reached. Common reasons are: too many samples, some grease on the hatch, a hair on the O-ring etc. If it is obvious what is the cause, it is also obvious what is the solution. Sometimes it is necessary to repeat an action and since it takes quite some time to get to the right vacuum level it is advised to pay special attention performing such an action; take away not one but a few samples if there are too many samples, clean the hatch a few times if that seems to be required and doublecheck the O-ring if you've discovered a hair.

The vacuum in the analysis chamber should be 5.10-8 mbar or less. If the vacuum is too high it could be resolved by moving the sample holder into the analysis chamber or by opening the bypass valve, valve V3. Valve V3 can't be controlled directly, but by turning on the flood gun it will be opened. Shutting down the flood gun again will keep valve V3 open!

If it seems that the vacuum gets better and better, then it may be a forewarning of a failing sensor. If the sensor of the loadlock fails, it can be replaced by the technician. If the sensor of the analysis chamber fails, it has to be replaced by the service engineer.


XPS info / ChemE / Delft University of Technology
version 20190503
anything else...
© 2000 - 2019 morechemistry.com. All rights reserved.
Design by Bb.