OUT_MODE_@0_# set mode of operation IN_MODE_@0 read mode of operation IN_PAR_@1 read controller resolution IN_PAR_@4 read kind of airflap control IN_PAR_@5 read kind of convection IN_PAR_@9 read 2nd temperature available IN_PAR_@A read pressure available IN_PV_@1 read actual temperature IN_PV_@2 read actual CO2 value IN_PV_@3 read actual rh value IN_PV_@5 read actual 2nd temperature IN_PV_@A read actual pressure IN_SP_@1 read actual temperature setpoint IN_SP_@4 read air flap position IN_SP_@5 read fan speed IN_SP_@A read vacuum setpoint OUT_SP_@1_-###.# write temperature setpoint OUT_SP_@4_# write air flap position OUT_SP_@5_# write fan speed OUT_SP_@6_# write volt free contact A OUT_SP_@7_# write volt free contact B OUT_SP_@8_# write volt free contact C OUT_SP@A_#### write Solwert pressure All commands must be followed by and first return OK @ is the address of the oven numeric formats: -###.# -999.9 .. 999.9 temperature ##.# 00.0 .. 99.9 CO2 value ## 00 .. 99 rh value #### 0000 .. 9999 pressure # 0 or 1 (with different meanings) # 0 .. A (0=off A = 100%)